TechTip: How do you know when a suspected problem really is a problem and not imagined?

Failing to identify a problem early can result in it possibly becoming a major problem with lots of down-time and costs involved in solving and fixing it. Many samples may also need to be re-analysed adding to costs, time, stress and customer satisfaction. But, thinking that there is a problem when there isn't really one can also lead to all of these things with lots of tail-chasing and not really getting anywhere.
 

Tech Tip: Rotary Pump Maintenance – Oil changes and Use of the Ballast Valve

Most GC-MS vacuum systems comprise of a two-stage pumping arrangement - an initial low-vacuum, high capacity stage consisting of a rotary oil pump or diaphragm pump; and a second high stage, low capacity vacuum pump such as a diffusion or turbo-molecular pump.  The first stage or low-vacuum pump requires maintenance to ensure optimum pumping performance.  Two aspects that are often overlooked are the use of the ballast valve on the vacuum pump and replacement of the pump oil at the correct intervals.

Tech Tip: Using a press-fit capillary column connector

Tech Tip: When to replace a gas cylinder

Knowing where your gas cylinders are located and regularly checking the contents is very important for any gas chromatographer when troubleshooting an instrument. People responsible for the cylinders go on holiday or leave and the first the instrument operators knows about it is when they run out of gas and then it is a major panic!
 
Unfortunately, many problems have already occurred by then.
 

Tech Tip: Random loss in sensitivity of injected samples through a batch

Most problems that occur in Gas Chromatography either gradually increase in severity over time or suddenly become a large issue, preventing further analysis by the instrument.

Random problems occurring infrequently are often ignored for long periods of time as the sample is simply re-analysed and the results are correct the second time. The problem itself can be hard to track down to determine the cause and then to provide a fix, as it isn't a continuous problem.

HOW TO FIX THIS

Tech Tip: Splitless injections - how long until you open the split exit?

Splitless injections are used to transfer all of the injected analytes onto the analytical column for separation and detection and are therefore used for low concentration samples. Opening the split exit too early causes any remaining analytes in the inlet to be flushed out of the split exit and lost. Opening it too late (or not at all) leads to a long, tailing solvent peak which can interfere with the separation and detection of early eluting analytes and provides a higher baseline well into the chromatogram.

Tech Tip: Separation - standard column or something more specialised?

 Obtaining the best resolution for the method is critical for accurate qualitative and quantitative analysis, especially where chromatographic resolution is relied upon with standard GC detectors, as there is no MSD to take advantage of analytical resolution.

 
 

Tech Tip: The Power of Sample Introduction

Sample introduction or the GC inlet parameters are very often overlooked when developing a GC or GC-MS method, but these parameters are so powerful that they can result in success or total failure of an application and as a result are also the main area for troubleshooting and maintenance within the GC system.

 

Tech Tip: Use your GC or GC-MS instrument to its full capabilities

Would you believe that only a small percentage of GC and GC-MS instruments are used to their full capabilities? This is the same for the data analysis systems.

In today's global climate of recession we really need to be making the most of what we have with optimised methods (both the analysis and the data processing) for high throughput and accurate analyses.

Ways to improve your methodologies: